TL;DR
France’s national blood service, the Établissement français du sang, says a cybersecurity incident at a refrigeration contractor affected cold-room control systems at three sites overnight on Nov. 1–2. Frozen plasma was moved as a precaution after alarms detected abnormal settings; EFS reports no health impact or loss of blood products. A reported 110°C value was a programmed target, not a measured room temperature.
France’s national blood service says a cybersecurity incident at a refrigeration contractor affected cold-room monitoring and control systems at three EFS sites overnight from Nov. 1 to Nov. 2, prompting staff to move frozen plasma as a precaution. The Établissement français du sang (EFS) says its alarms detected the abnormal settings, with no reported health impact or loss of blood products.
In a statement issued Nov. 3 and reported by franceinfo and Next on Nov. 4, EFS said a contractor responsible for refrigeration equipment had suffered a cybersecurity incident. It said the incident affected supervision and control installations for cold rooms at three of its sites, where temperature and defrost thresholds were raised to very high levels. EFS did not identify the contractor or publish the altered settings.
Le Canard enchaîné reported that the rooms were set to run at −30°C and that a controller was changed to a target of +110°C. That figure is a programmed setpoint, not a measured temperature: no source cited in the reporting says any room reached 110°C. The EFS statement itself did not specify a temperature.
EFS said its monitoring and alarm systems identified the abnormal settings quickly. The equipment was brought under control or switched off, and the contents were moved to other facilities as a precaution. The rooms held frozen plasma for transfusions and plasma intended for fractionation, a process used to make medicines. EFS reported no loss of blood products and no identified health impact. Its technical teams are working with France’s national cybersecurity agency, ANSSI, and the health ministry, while monitoring has been increased across EFS equipment.
Infographic 01 · 110 °C, explained
A setting, not a temperature
Someone raised the targets of plasma cold rooms at France’s blood service to +110 °C. The rooms never got there. What mattered was the alarm.
Frozen plasma has to stay at or below this line. The danger was thawing, not boiling. A few degrees too warm can already mean discarding bags.
- +110 °CThe programmed setpointTyped into the controller (Canard). Never reached, never measured.
- ~+50 °CHottest reported in the supermarket waveDefrost heaters forced on, same night (AFP, ZDFheute)
- −25 °CThe line that mattersFrozen plasma is stored at −25 °C or colder (EDQM guide via ISBT)
- −30 °CNormal operation of the EFS cold roomsAs reported by Le Canard enchaîné
Why the alarm won
- DetectionEFS monitoring and alarms flagged the anomaly quickly (EFS)
- A person on callAn on-call technician saw the temperature rise in the night (Canard)
- A plan BThe plasma was moved to other installations as a precaution (EFS)
Sources: EFS communiqué, 3 Nov 2026, via franceinfo · Le Canard enchaîné, as reported by Solutions Numériques · ISBT, citing the EDQM Guide · AFP via Al Jazeera, 4 Nov · ZDFheute, 3 Nov. The 110 °C and −30 °C figures come only from Le Canard enchaîné; the EFS gives no number.
Powered by Thorsten Meyer AI
Why Plasma Cold Storage Matters
Frozen plasma must remain within strict temperature limits to preserve its suitability for medical use. The International Society of Blood Transfusion, summarizing European guidance, says fresh frozen plasma is stored at −25°C or colder. A sustained rise above the required range could put stock at risk, although EFS says the products in this incident were moved and reports no loss.
The incident also shows why independent alarms and an on-call response matter in systems connected to remote maintenance. Here, EFS says its monitoring detected the abnormal settings and staff relocated the plasma. That response prevented a reported product loss, but the available information does not establish how long the systems were exposed or what temperatures the rooms actually reached.
The event occurred on the same night as a second wave of attacks affecting supermarket refrigeration in several European countries, according to the source reporting. The timing raises a question about exposure across cold-storage operations, but no authority has linked the incidents. They should not be treated as one coordinated campaign on the available evidence.
Popular right now

Philips Café Aromis Automatic Espresso Machine with LatteGo Pro Milk System

Ninja Foodi Dual Zone Smart XL Air Fryer, 10 Qt, 6-in-1, Black (DZ550)

Hisense 65-inch U6 Pro Mini-LED 4K UHD Smart Fire TV
As an affiliate, we earn on qualifying purchases.
How the EFS Incident Emerged
EFS is France’s national blood service, collecting donations and supplying hospitals. The incident was first detailed publicly by Le Canard enchaîné in its Nov. 3 issue. The EFS statement, as reported by franceinfo and Next, confirmed an incident at a refrigeration contractor and the effect on control installations at three sites, but did not name the contractor.
Le Canard enchaîné identified the contractor as Coolworld, which rents and maintains refrigerated rooms and containers. That identification is based on the newspaper’s reporting; the source material says Coolworld had not commented publicly, and EFS said it was not directly targeted. Bourgogne-Franche-Comté was identified as one affected region, while the other two sites were not made public.
Reports also described attacks on supermarket refrigeration systems during the same night, including altered settings that triggered defrost heating. Two groups describing themselves as pro-Russian later claimed the supermarket attacks in a video, according to the source material. The claims were unverified, the posts were deleted, and no government had publicly attributed those attacks to a group. No connection to the EFS incident has been established.
““The EFS monitoring and alarm systems made it possible to identify these anomalies quickly.””
— Établissement français du sang
What the Public Record Leaves Open
EFS has not named the refrigeration contractor or disclosed the two other affected sites. Its statement also does not provide the exact altered settings, the measured temperatures, the duration of the incident, or technical details about how access was obtained. The +110°C figure comes from Le Canard enchaîné, not the EFS communiqué, and refers to a controller target rather than a recorded temperature.
The contractor’s public response was not available in the source material. It is also not clear whether the incident affected any other contractor customers or systems. Authorities have not publicly identified who was responsible, and there is no confirmed link between this event and the contemporaneous supermarket refrigeration attacks or the groups that claimed those attacks.
EFS and Agencies Continue Review
EFS says technical teams are working with ANSSI and the health ministry to address the incident. It has also stepped up temperature monitoring across its equipment. The source material does not give a timetable for the technical review, identify any planned public report, or say whether the contractor’s remote access arrangements have been changed.
Further confirmed details may clarify how the control systems were accessed, what temperatures were recorded, and whether investigators establish a link to any other incident. For now, EFS’s reported outcome is that the plasma was moved, no blood products were lost, and no health impact has been identified.
Key Questions
Did a plasma freezer actually reach 110°C?
No measured temperature of 110°C has been reported. Le Canard enchaîné reported that 110°C was entered as a controller setpoint. EFS’s statement described very high settings but did not publish a number.
Were any plasma products lost?
EFS says no blood products were lost. It reported that the contents were moved to other installations as a precaution and that no health impact had been identified.
Where did the incident happen?
EFS said cold-room control installations at three sites were affected. Le Canard enchaîné identified Bourgogne-Franche-Comté as one region; the other two sites have not been made public.
Is this connected to the attacks on supermarket refrigeration?
No connection has been established. The incidents happened on the same night, but the source material says no authority has linked them. Claims by groups about the supermarket attacks were unverified and do not establish responsibility for the EFS incident.